1. Introduction
Cystic echinococcosis (CE) is a global zoonotic disease caused by the larval form of Echinococcus granulosus sensu lato ( 1 ). This parasite is mainly transmitted between dogs as the definitive hosts, while various livestock species (sheep, cattle, goats, and camels) serve as intermediate hosts. Humans serve as accidental intermediate hosts of this parasite. Infection in humans and intermediate hosts occurs through the accidental intake of E. granulosus eggs, usually present in the feces of infected canines, leading to CE infection ( 2 ). The disease is prevalent in various regions globally, particularly in the Mediterranean, Middle East, South America, Central Asia, Australia, Chile, Eastern Europe, New Zealand, and West Africa ( 3 ). Iran is regarded as an endemic area for E. granulosus sensu lato, encompassing different species and genotypes ( 4 ).
The disease causes substantial economic losses for both livestock farming and human health. In Iran, the average annual financial impact of CE is estimated to exceed $232.3 million ( 5 ). The disability-adjusted life years (DALY) of CE in 2018 were estimated to be 1,210 years in Iran (700 years in men and 510 years in women) ( 6 ). Timely diagnosis of this disease is very important in humans. CE is asymptomatic in most cases (about 65%) and is occasionally discovered following an examination for the diagnosis of other diseases ( 7 ). Achieving a conclusive diagnosis requires the integration of multiple methods, and careful consideration of clinical symptoms, epidemiological data, and the patient's medical history is essential.
Imaging methods such as CT scans, ultrasound, MRI, and X-ray radiography, along with immunological tests like ELISA, Western blot, indirect hemagglutination, indirect immunofluorescence, latex agglutination tests, and counter-current immunoelectrophoresis (CCIEP), constitute the primary methods used for diagnosing CE ( 4 , 8 ). The World Health Organization endorses the use of the international ultrasonography classification for staging CE through stage-specific imaging ( 9 ).
During recent years, different native and recombinant antigens with various sensitivity and specificity have been utilized for setting up immunological tests in definitive or intermediate hosts ( 10 - 13 ). Hydatid cyst fluid (HCF) is the most common antigen used in this field. The sensitivity of this antigen can vary between 50 and 98%, depending on factors such as the specific serological method used, the quality and purity of the antigen preparation, the genotype of the parasite, the stage and location of the cyst, the number and size of cysts, the patient's immunoglobulins (isotypes), and the parasite genotype ( 10 , 11 , 13 ).
Antigen B (AgB) that found in cyst fluid is a multimeric protein consisting of subunits with varying molecular weights (8, 16, 24, and 32 kDa). This protein exhibits high immunogenic properties, making it valuable for the diagnosis of human CE through ELISA tests. The predominant antibody response to the main reactive subunits, EgAgB1, EgAgB2, and EgAgB4, was found previously to be of the IgG4 subclass ( 14 ). The present study aimed to produce an in-house antigen B (AgB) and compare its diagnostic efficacy with that of a domestic commercial kit (Pishtaz Teb Company, Iran), using ELISA method.
2. Materials and Methods
A total of 142 serum samples were classified into three groups: the first group included 36 sera from patients with CE confirmed by pathology (as positive controls), the second group included 60 sera from healthy people (as negative controls), and the third group consisted of 46 sera from individuals with other parasitic diseases including giardiasis (n=6), toxoplasmosis (n=2), malaria (n=5), leishmaniasis (n=1), taeniasis (n=3), fascioliasis (n=6), toxocariasis (n=5), hymenolepiasis (n=6), ascariasis (n=5), trichostrongyliasis (n=1), strongyloidiasis (n=1), enterobiasis (n=1), blastocystosis (n=3), and amoebiasis (n=1).
2.1. Preparation of HCF
HCF was collected from fertile cystic echinococcosis cysts in sheep in the parasitology laboratory of Baqiytallah University of Medical Sciences. To remove the protoscoleces and large particles, HCF was subjected to centrifugation at 3000×g for 15 minutes at 4 ºC and kept at −20 ºC until use.
2.2. Preparation of AgB
The purification of AgB from HCF was carried out based on the established protocol by Oriol ( 15 ). In this regard, 100 mL of HCF underwent overnight dialysis at 4°C using 5 mM acetate buffer (pH 5). The dialyzed content was centrifuged at 50,000×g for 30 min. The supernatant was discarded, and the resulting pellet was reconstituted in 0.2 M phosphate buffer (pH 8). To eliminate globulins, saturated ammonium sulfate precipitation was employed. Subsequent heat treatment involved boiling the sample for 15 min, followed by another centrifugation step (50,000×g, 60 min) to isolate the thermostable AgB fraction in the supernatant. Protein quantification was performed via the Bradford assay, in which BSA served as the reference standard ( 16 ).
2.3. Evaluation of sera by ELISA
Microplate-based immunoassays were performed using 96-well flat-bottom plates. Each well was coated with 100 µL of AgB (5 µg/mL in 0.05 M carbonate-bicarbonate coating buffer, pH 9.6), and the plate was incubated overnight at 4°C. Unbound antigens were eliminated through successive washes with PBS-T (0.05% Tween 20 in phosphate-buffered saline, pH 7.4). Non-specific binding sites were blocked by 100 µL of 3% skimmed milk (Merck, Germany) in PBST, incubated for 2 hours at room temperature. Following three washes with PBST, the plates were incubated with 100 µL of serum samples (diluted 1:100 in PBST) for 90 minutes at room temperature. After five washes, 100 µL of HRP-conjugated secondary antibody (Sigma, USA; 1:4000 dilution in PBST) was added, and the plate was incubated for 1 hour at room temperature.
Following a final wash cycle as above, 100 µL of OPD substrate solution (0.025% o-phenylenediamine dihydrochloride with 0.1% H2O2 in 0.1 M citrate buffer, pH 5.0; Sigma, USA) was added to each well, and the plate was incubated in the dark for 30 minutes. The enzymatic reaction was stopped by adding 1 M sulfuric acid. Absorbance was measured at 492 nm using a BIOTEK ELx800 microplate reader. The diagnostic cutoff was established as the mean optical density of negative controls plus two standard deviations.
2.4. Evaluation of sera using commercial ELISA Kit
The IgG antibody produced against Echinococcus antigens in human sera was measured using the Echinococcus IgG ELISA kit (Pishtaz Teb Co., Iran). In each flat-bottom 96-well microplate, four wells were designated as negative control (two wells), positive control, and the blank. Other wells received 100 μL of diluted serum samples (1:100 in PBS), and the plate was incubated at room temperature for 30 minutes. After five washes, horseradish peroxidase-conjugated secondary antibody (100 μL/well) was added to the wells (excluding blanks), and the plate was incubated at room temperature for 30 minutes. After repeating the washing procedure as above 100 μL of chromogenic substrate (TMB or equivalent) was added to each well and developed in the dark for 15 minutes at room temperature.
The reaction was stopped with 100 μL of 2 M sulfuric acid, and absorbance was immediately measured at 496 nm using a BIOTEK ELx800 microplate reader. The diagnostic cut-off was calculated by adding 0.25 to the average OD of negative controls in each microplate. Samples with an OD 10% higher than the cut-off were considered positive.
2.5. Data analysis
The validation parameters of AgB andthe commercial ELISA Kit were determined using MEDCALC software ( 17 ).
3. Results
The AgB-based indirect ELISA demonstrated high diagnostic performance, with 86.11% sensitivity and 95.28% specificity for detecting human anti-CE antibodies. This antigen showed false-positive results in two patients with fascioliasis and three patients with taeniasis, ascariasis, and toxocariasis infections.
Evaluation revealed high diagnostic performance for AgB, showing 92.96% efficiency and 90.6% validity in CE antibody detection. The commercial kit presented a lower number of false-positive results in other parasitic sera (one serum corresponding to taeniasis) than AgB (five sera). The diagnostic efficiency and validity of AgB were determined to be 92.96% and 90.69%, respectively. For the commercial ELISA kit, these values were reported as 93.66% and 88.42%, respectively. The diagnostic performance of our AgB-based ELISA was systematically compared with a commercial CE detection kit through validation parameters (Table 1).
| Statistical index | No. of samples | No. of seropositive samples | |
|---|---|---|---|
| AgB | Commercial kit | ||
| CE samples | 36 | 31 | 28 |
| Samples from healthy people | 60 | 0 | 0 |
| Samples from patients with other parasitic diseases | 46 | 5 | 1 |
| Sensitivity | 86.11 | 77.78 | |
| Specificity | 95.28 | 99.06 | |
| Positive predictive value | 86.11 | 96.55 | |
| Negative predictive value | 95.28 | 92.92 | |
| Diagnostic efficiency | 92.96 | 93.66 | |
| Validity | 90.69 | 88.42 | |
4. Discussion
CE represents a clinically and economically significant zoonosis affecting both human populations and livestock, particularly in endemic regions of the Middle East, including Iran ( 2 , 3 ). This disease is of special importance from medical, veterinary, and economic points of view. CE causes significant economic impacts on both livestock husbandry and human populations due to the direct and indirect costs incurred from medical treatments and losses related to livestock ( 5 ).
Imaging methods and serological testing are used together for the diagnosis of human diseases, providing a more comprehensive assessment of the patient's condition.
Immunological diagnosis, as well as new methods, can be applied not only for initial diagnosis but also for follow-up of patients after treatment ( 18 ). Among these, ELISA is widely used by researchers as a standard method due to its remarkable sensitivity and specificity in the diagnosis of human CE ( 19 ). One of the main pitfalls of serological tests is cross-reactivity with other parasitic diseases. The cross-reactivity of antigens applied for the serodiagnosis of CE with parasites of the Taeniidae family (Echinococcus multilocularis, Taenia solium, etc.) is a drawback for the diagnosis of CE, as they have a close antigenic affinity with E. granulosus, leading to a reduction in the specificity of serological tests ( 20 ). HCF is the common component for the diagnosis of CE in humans as a crude antigen. AgB and Ag5 are the main antigens present in HCF. AgB, a 120–160 kDa oligomeric thermostable lipoprotein composed of three subunits at 8/12, 16, and 20 kDa, is widely used for the serodiagnosis of human CE ( 11 , 21 ).
Although recombinant proteins have received attention for the serodiagnosis of human CE, the use of recombinant antigens presents challenges such as antigen storage, advanced laboratory equipment requirements, and standardization, etc. ( 10 ). Simplicity, easy access, acceptable accuracy, and cost-effectiveness are crucial factors to consider when choosing a diagnostic method. In this study, in-house AgB was compared with the commercially available ELISA kit (Pishtaz Teb Co., Iran). One hundred forty-two serum samples (36 pathologically confirmed sera from patients with CE, 46 sera from individuals with other parasitic diseases, and 60 sera from healthy individuals) were evaluated by indirect ELISA, where a sensitivity of 86.11% and a specificity of 95.28% were obtained for the diagnosis of anti-CE antibodies in humans.
The commercial kit showed a sensitivity of 77.78% and a specificity of 99.06% in detecting anti-CE antibodies in humans. Also, AgB and the commercial kit showed diagnostic efficiencies of 92.96% and 93.66%, respectively. AgB and commercial kit showed the validity of 90.69% and 88.42%, respectively, for the detection of anti-CE antibodies in humans.
Lorenzo et al. examined a panel of 129 sera, consisting of 59 sera from patients with CE, 55 sera from patients with other diseases, and 15 sera from healthy individuals using the AgB-ELISA test. The sensitivity and specificity in this study were determined to be 80% and 77%, respectively. Sera from individuals with neurocysticercosis and alveolar echinococcosis were reported as cases of cross-reactivity ( 22 ). The antigenic diversity of hydatid cyst fluid, differences in parasite genotype, the duration of the disease, and the time after treatment are among the factors involved in negative or false-positive results in serology tests ( 23 ).
Another study conducted by Mohammadzadeh et al. ( 11 ) examined three types of antigens including AgB (AgB taken from infected sheep in Iran and China), a recombinant AgB, and HCF, using the ELISA method for serodiagnosis of CE. The sensitivities of Iranian and Chinese AgB, recombinant antigen, and cyst fluid were 96.4%, 82.1%, 94.6%, and 91.1%, respectively, and their specificities were 84.8%, 94.9%, and 93.9% and 76.8%, respectively ( 11 ).
Reiterova et al. ( 24 ) compared AgB-ELISA and FH-ELISA produced from sheep hydatid cysts. For this purpose, 177 sera, including 50 negative control samples, 55 positive control samples, and 72 samples from other diseases, were used. The sensitivity and specificity of the AgB-ELISA were reported as 96.4% and 97.2%, respectively, where the sensitivity of FH-ELISA was higher than that of AgB-ELISA, while the specificity of AgB-ELISA was higher than that of FH-ELISA ( 24 ). Evaluation of four distinct antigen preparations derived from Echinococcus granulosus sensu stricto —including native HCF, lyophilized HCF, AgB, and lyophilized AgB —for serological diagnosis of active, transitional, and inactive hepatic CE demonstrated that the anti-LHCF IgG ELISA achieved the highest sensitivity (96.97%), while the LAgB-based ELISA showed optimal specificity (95.37%). These findings indicate that a combined diagnostic approach using both anti-LHCF IgG ELISA and anti-LAgB IgG ELISA would provide the most accurate serodiagnosis of human liver CE ( 25 ).
Despite the use of the same antigen in some studies, the efficacy of the test is not completely the same. The number and type of sera used, especially sera from individuals with other diseases, are additional reasons for these differences. In many studies, the stage or condition of the cyst in patients is not taken into account during evaluations. Also, collecting sera over a long period can affect the quality of serum antibodies. This study showed the very favorable sensitivity, specificity, and validity of AgB in the serodiagnosis of human CE. The sensitivity of AgB in the present research was higher than that of commercial kit (86.11% vs. 77.78%), and its specificity was found to be lower than that of the aforementioned kit (95.28% vs. 99.06%). Overall, the validity of in-house AgB in the present research was higher than that of the commercial kit (90.69% vs. 88.42%), and the diagnostic efficiency of both antigens showed a slight difference of less than 1%.
Sera from people suffering from other parasitic diseases that responded positively with AgB (cross-reaction) included two cases of fascioliasis, and one case each of taeniasis, ascariasis, and toxocariasis. There is a possibility of simultaneous infection with Fasciola spp., Taenia spp., and Ascaris lumbricoides parasites due to the transmission cycles of these parasites through eating contaminated vegetables.
The results of various studies show a wide range of sensitivity and specificity, and these differences in each study can be affected by factors such as the type and structure of the antigen used, the number of sera evaluated, the sampling method and their maintenance, cyst stage, type of diagnostic techniques, base test for comparison, and determination of validation parameters, etc. In conclusion, although both in-house-produced AgB and the commercial kit showed considerable and relatively similar diagnostic efficiency, the study results demonstrated higher sensitivity and validity for AgB compared with the commercial ELISA kit.
Acknowledgment
This study was conducted as part of the MD thesis by author Saidae Arsalani, under the ethical code IR.BMSU.REC.1398.224, at the Department of Parasitology and Mycology, School of Medicine, Baqiyatallah University of Medical Sciences. It was supported by the Research Deputy of the School of Medicine.
Authors' Contribution
Study concept and design: T. M.
Acquisition of data: S.A, F.H, PS, E.D
Analysis and interpretation of data: T.M, F.H
Drafting of the manuscript: E.D, S.H
Critical revision of the manuscript for important intellectual content: S.M.S
Statistical analysis: T.M, F. H
Administrative, technical, and material support: T.M
Ethics
Ethical code: IR. BMSU.1398.224.
Conflict of Interest
The authors declare that there is no conflict of interest.
Data Availability
All data generated are included in the current article.
References
- Thompson R. The taxonomy, phylogeny and transmission of Echinococcus. Exp Parasitol. 2008; 119(4):439-46.
- Eckert J, Deplazes P. Biological, epidemiological, and clinical aspects of echinococcosis, a zoonosis of increasing concern. Clin Microbiol Rev. 2004; 17(1):107-35.
- Sadjjadi SM. Present situation of echinococcosis in the Middle East and Arabic North Africa. Parasitol Int. 2006; 55:S197-202.
- Borhani M, Fathi S, Darabi E, Jalousian F, Simsek S, Ahmed H, et al. Echinococcoses in Iran, Turkey, and Pakistan: old diseases in the new millennium. Clin Microbiol Rev. 2021; 34(3):e00290-20.
- Fasihi Harandi M, Budke CM, Rostami S. The monetary burden of cystic echinococcosis in Iran. PLoS Negl Trop Dis. 2012; 6 (11):e1915.
- Parandin F, Heydarpour F, Mohebali M, Hanafi-Bojd AA, Sari AA, Zeynali M, et al. Estimation of Burden of Cystic Echinococcosis in Iran Using Disability Adjusted Life Years (DALYs) in 2018. Iran J Public Health. 2021; 50 (11):2302.
- Eckert J, Gemmell M, Meslin FX, Pawlowski Z, World Health Organization. WHO/OIE manual on echinococcosis in humans and animals: a public health problem of global concern. World Organisation for Animal Health; 2001.
- Torgerson PR, Deplazes P. Echinococcosis: diagnosis and diagnostic interpretation in population studies. Trends Parasitol. 2009; 25(4):164-70.
- WHO Informal Working Group. International classification of ultrasound images in cystic echinococcosis for application in clinical and field epidemiological settings. Acta Trop. 2003; 85(2):253-61.
- Darabi E, Motevaseli E, Mohebali M, Rokni MB, Khorramizadeh MR, Zahabiun F, et al. Evaluation of a novel Echinococcus granulosus recombinant fusion B-EpC1 antigen for the diagnosis of human cystic echinococcosis using indirect ELISA in comparison with a commercial diagnostic ELISA kit. Exp Parasitol. 2022; 240:108339.
- Mohammadzadeh T, Sako Y, Sadjjadi SM, Sarkari B, Ito A. Comparison of the usefulness of hydatid cyst fluid, native antigen B and recombinant antigen B8/1 for serological diagnosis of cystic echinococcosis. Trans R Soc Trop Med Hyg. 2012; 106(6):371-5.
- Jalousian F, Hosseini SH, Fathi S, Shirani D, Aghaei S, Kordafshari S. Comparative assessment of rEPC1 antigen and copro-antigen for diagnosis of echinococcosis in dogs. Iran J Vet Med. 2017; 11(3):217-225.
- Farhoudi R, Hosseini SH, Siavashi MR, Abolhassani M, Shayan P, Nejadmoghadam A. Identification and purification of a specific and immunogenic antigen of the laminated layer of the hydatid cyst and production of an antigen-specific monoclonal antibody. Int J Vet Res. 2010; 4(4): 237-43.
- Lightowlers MW, Liu D, Haralambous A, Rickard MD. Subunit composition and specificity of the major cyst fluid antigens of Echinococcus granulosus. Mol Biochem Parasitol. 1989; 37(2):171-82.
- Oriol R, Williams J, Pérez E, Oriol C. Purification of lipoprotein antigens of Echinococcus granulosus from sheep hydatid fluid. Am J Trop Med Hyg. 1971; 20(4):569-74.
- Bradford MM. A rapid and sensitive method for the quantitation of microgram quantities of protein utilizing the principle of protein-dye binding. Anal Biochem. 1976; 72(1–2):248-54.
- Schoonjans F, Zalata A, Depuydt CE, Comhaire FH. MedCalc: a new computer program for medical statistics. Comput Methods Programs Biomed. 1995; 48(3):257-62.
- Zhang W, Wen H, Li J, Lin R, McManus DP. Immunology and immunodiagnosis of cystic echinococcosis: an update. Clin Devel Immunol. 2012; 2012:1-10.
- Virginio V, Hernandez A, Rott M, Monteiro K, Zandonai A, Nieto A, et al. A set of recombinant antigens from Echinococcus granulosus with potential for use in the immunodiagnosis of human cystic hydatid disease. Clin Exp Immunol. 2003; 132(2):309-15.
- Schantz PM, Shanks D, Wilson M. Serologic cross-reactions with sera from patients with echinococcosis and cysticercosis. Am J Trop Med Hyg. 1980; 29(4):609-12.
- Barbieri M, Fernández V, González G, Luaces VM, Nieto A. Diagnostic evaluation of a synthetic peptide derived from a novel antigen B subunit as related to other available peptides and native antigens used for serology of cystic hydatidosis. Parasite Immunol. 1998; 20(2):51-61.
- Lorenzo C, Ferreira HB, Monteiro KM, Rosenzvit M, Kamenetzky L, García HH, et al. Comparative analysis of the diagnostic performance of six major Echinococcus granulosus antigens assessed in a double-blind, randomized multicenter study. J Clin Microbiol. 2005; 43(6):2764-70.
- Moro P, Schantz PM. Echinococcosis: a review. Intl J Infect Dis. 2009; 13(2):125-133.
- Reiterová K, Auer H., Altintaś N, Yolasigmaz A. Evaluation of purified antigen fraction in the immunodiagnosis of cystic echinococcosis. Parasit Res. 2014; 113:2861-7.
- Sadjjadi FS, Mohammadzadeh T, Jafari SH, Sharifi Y, Deilami HN, Hafezi F, Sadjjadi SM. Comparison of Native Hydatid Cyst Fluid (HCF), Lyophilized HCF, Antigen B (AgB) and Lyophilized AgB (LAgB) Originated from Echinococcus granulosus Sensu Stricto for Sero- Diagnosis of Active, Transitional and Inactive Human Liver Cystic Echinococcosis. Iran J Public Health. 2023; 52(8):1764-72.