1. Introduction
The element selenium, CAS No. 7782-49-2, is a member of the group VIA of the periodic table and possesses metallic and nonmetallic characteristics ( 1 ). Selenium belongs to class 2B of elements according to the International Council for Harmonization Q3D guideline ( 2 ). These class-2B elements are less likely to be included in the therapeutic product due to their limited availability and poor ability to co-isolate with other materials. Therefore, they might not be considered in the risk assessment unless they are consciously incorporated while producing drug compounds, excipients, or other components of the drug product ( 2 ). Selenium sulfide is an anti-infective agent with antifungal and antibacterial properties and is used to cure dandruff and seborrheic dermatitis ( 3 ); hence, it is used as a dermatological shampoo.
Literature survey shows that selenium can be determined using different techniques, such as voltammetric analysis ( 4 , 5 ), spectrophotometry ( 6 , 7 ), chromatography (high-performance liquid chromatography [HPLC]) ( 8 , 9 ), atomic absorption spectroscopy ( 10 - 14 ), atomic fluorescence spectroscopy (AFS) ( 15 - 17 ), inductively coupled plasma optical emission spectroscopy (ICP-OES) ( 18 ), inductively coupled plasma mass spectrometry (ICP-MS) ( 19 ), HPLC combined with ICP-MS ( 20 ), and HPLC in conjunction with AFS ( 21 ). Most of the above-mentioned analytical techniques were found to be time-consuming and/or required high levels of skills. Inductively coupled plasma optical emission spectroscopy is the preferred technique as it is cost-effective for single-element analysis and is sensitive and reliable. On the other hand, the open digestion technique is dangerous, time-consuming, and inconsistent. Open digestion sample preparation followed by a non-specific titration was used for estimating selenium sulfide drug substance ( 22 ) and selenium sulfide topical suspension ( 23 ). This research work aimed to develop and validate a simple, cost-effective, robust ICP-OES method for estimating selenium in selenium sulfide drug substance and shampoo formulation using closed microwave digestion.
2. Materials and Methods
2.1. Chemicals, reagents, standards, and samples
Selenium sulfide drug substance was purchased from ABCR GMBH (Germany). Selsun blue dandruff shampoo with menthol, 1% selenium sulfide (Manufacturer: Sanofi), and selenium sulfide topical suspension USP, 2.5% lotion (Manufacturer: Perrigo) were bought from the open market. Selenium 10 mg/ml was acquired from Inorganic Ventures. All the chemicals used were of analytical reagent grade. Ammonium hydrogen difluoride was bought from Sigma-Aldrich (USA). Boric acid 99.99% was obtained from Alfa Aesar (USA). Concentrated nitric acid (70%) trace metal grade was acquired from Fisher (USA).
2.2. Instruments and equipment:
ICP-OES made of Thermo-scientific iCAP 6000 series with Helium/Kinetic energy discrimination mode was utilized. The microwave digestion system CEM MARS 6 with Easy Prep Plus Vessels was used. The Mettler Toledo analytical balance with model: ME204E was employed. A hot plate made by Royal Scientific (model RSW 127) was used.
2.3. Sample preparation:
The first phase in the technique development process involved the investigation of solubility. The method development trail was initiated by dissolving approximately 0.1 g of drug substance (i.e., selenium sulfide) in 100 ml of water. Similarly, 0.1 g of shampoo and topical suspension were dissolved in 100 ml of water. As an alternate method, closed microwave digestion with concentrated high-pure acids was considered for sample preparation. Closed vessel microwave digestion was initiated by using a varied ratio of acids. The trial observations are tabulated in table 1. All the sample matrices were subjected to microwave-assisted closed vessel digestion under controlled pressure and temperature.
S. No. | Selenium Sulfide Topical Suspension | Closed vessel microwave-assisted digestion program | Selenium Sulfide drug substance | Selsun Blue Dandruff Shampoo | Selenium Sulfide Topical Suspension | |||
---|---|---|---|---|---|---|---|---|
1 | Samples + 5ml conc. Nitric acid + 5ml conc. hydrochloric MWD | Step | Ramp (min) | Hold (min) | Temp. (°C) | The sample settles in the bottom of the vessel with partial digestion | The sample settles in the bottom of the vessel with partial digestion | The sample settles at the bottom of the vessel with partial digestion |
1 | 20 | 30 | 200 | |||||
2 | Samples + 5ml conc. Nitric acid + 5ml conc. hydrochloric acid MWD Step-1 + 10ml conc. Nitric acid MWD Step-2 | Step | Ramp (min) | Hold (min) | Temp. (°C) | The sample settles in the bottom of the vessel with partial digestion | The sample settles in the bottom of the vessel with partial digestion | The sample settles at the bottom of the vessel with partial digestion |
1 | 20 | 30 | 200 | |||||
2 | 20 | 10 | 180 | |||||
3 | Samples + 0.2g ammonium hydrogen difluoride + 1ml water predigested for 5 minutes + 5ml conc. Nitric acid predigested for 15 minutes MWD Step-1 + 5ml 4% boric acid MWD Step-2 | Step | Ramp (min) | Hold (min) | Temp. (°C) | The sample settles in the bottom of the vessel with partial digestion | The sample settles in the bottom of the vessel with partial digestion | The sample settles in the bottom of the vessel with partial digestion |
1 | 10 | 15 | 180 | |||||
4 | Samples + 0.2g ammonium hydrogen difluoride + 1ml water predigested for 5 minutes + 5ml conc. Nitric acid predigested for 15 minutes MWD Step-1 + 5ml 4% boric acid MWD Step-2 | Step | Ramp (min) | Hold (min) | Temp. (°C) | The sample settles in the bottom of the vessel with partial digestion | The sample settles in the bottom of the vessel with partial digestion | The sample settles in the bottom of the vessel with partial digestion |
1 | 10 | 15 | 200 | |||||
5 | Samples + 0.2g ammonium hydrogen difluoride + 1ml water predigested for 5 minutes + 5ml conc. Nitric acid predigested for 15 minutes MWD Step-1 + 5ml 4% boric acid MWD Step-2 | Step | Ramp (min) | Hold (min) | Temp. (°C) | Almost all the sample digests with some residue at the bottom of the vessel | Almost all the sample digests with some residue at the bottom of the vessel | Almost all the sample digests with some residue at the bottom of the vessel |
1 | 20 | 30 | 200 | |||||
2 | 20 | 10 | 180 | |||||
6 | Samples + 0.5g ammonium hydrogen difluoride + 1ml water predigested for 5 minutes + 7ml conc. Nitric acid predigested for 15 minutes MWD Step-1 + 7ml 4% boric acid MWD Step-2 | Step | Ramp (min) | Hold (min) | Temp. (°C) | Complete digestion | Almost all the sample digests with some residue at the bottom of the vessel | Almost all the sample digests with some residue at the bottom of the vessel |
1 | 20 | 30 | 200 | |||||
2 | 20 | 10 | 180 | |||||
7 | Samples + 0.7g ammonium hydrogen difluoride + 1ml water à predigested for 5 minutes + 10ml conc. Nitric acid à predigested for 15 minutes MWD Step-1 + 10ml 4% boric acid MWD Step-2 | Step | Ramp (min) | Hold (min) | Temp. (°C) | Complete digestion | Complete digestion | Complete digestion |
1 | 20 | 30 | 200 | |||||
2 | 20 | 10 | 180 |
Sample preparation optimization
Selenium sulfide drug substance
For the drug substance, 0.5 g of selenium sulfide was accurately weighed. Approximately 0.7 g of ammonium hydrogen difluoride was added to each vessel, followed by 1 ml of water. The samples were pre-digested for approximately 5 min in the ammonium hydrogen difluoride solution. A volume of 10 ml of concentrated nitric acid was added to each digestion vessel, and the samples were continued to pre-digest for around 15 min. The microwave reaction vessels were assembled, and the samples were digested according to the parameters described in the digestion step 1 (Table 3). The samples were allowed to cool before adding 10 ml of 4% boric acid required to complex the remaining hydrofluoric acid. The microwave reaction vessels were reassembled, and the samples were digested according to the parameters described in the digestion step 2 (Table 3). Digested samples were quantitatively transferred to individual 10 ml volumetric flasks. Samples were allowed to cool to room temperature before diluting to volume with water.
Parameters | Settings | Digestion Step 2 Settings |
---|---|---|
Power | 60 – 1800 watts | 400 – 1800 watts |
Ramp Time | 20 mins | 20 mins |
Hold Time | 30 mins | 10 mins |
Temperature | 200 °C | 180 °C |
Pressure Max | 800 psi | 800 psi |
Cool Down Time | 30 mins | 30 mins |
Selenium sulfide drug formulation
Samples were prepared by shaking the product container vigorously and mixing a portion of the shaken product in a beaker to ensure sample homogeneity. A volume of 0.5 g of Selsun blue dandruff shampoo and 0.2 g of selenium sulfide topical suspension USP were accurately weighed separately. Using a micropipette pipette, samples were dispensed into tared microwave vessels, and the product weight was recorded. Approximately 0.7 g of ammonium hydrogen difluoride was added to each vessel, followed by 1 ml of water. The samples were pre-digested for approximately 5 min in the ammonium hydrogen difluoride solution. Subsequently, 10 ml of concentrated nitric acid was added to each digestion vessel, and the samples were continued to pre-digest for around 15 min. The microwave reaction vessels were assembled, and the samples were digested according to the parameters described in the digestion step 1 (Table 3). The samples were allowed to cool before adding 10 ml of 4% boric acid required to complex the remaining hydrofluoric acid. The microwave reaction vessels were reassembled, and the samples were digested according to the parameters described in the digestion step 2 (Table 3). Digested samples were quantitatively transferred to individual 100 ml volumetric flasks. Samples were allowed to cool to room temperature before diluting to volume with water.
2.4. Rinsing solution for ICP-OES instrument autosampler - 2% nitric acid (v/v):
Slowly, 20 ml of nitric acid (concentrated) was added to 500 ml of water. This solution was diluted to 1 L with water and mixed well.
2.5. Preparation of 4% boric acid (w/v):
Approximately 40 g of boric acid was weighed and added to 1,000 ml of water. The mixture was stirred while gently heated until it completely dissolved.
2.6. Preparation of standard solutions:
Concentrated nitric acid, 4% boric acid, and 10 mg/ml selenium (Se) reference standard were transferred in 500 ml and 200 ml volumetric flasks accordingly (as volumes indicated in Tables 2 and 3) and diluted to volume with water to prepare calibration standards and calibration check standards.
Standard Concentration (μg/ml) or (ppm) | Volume (ml) Concentrated Nitric Acid | Volume (ml) 4% Boric Acid | Volume (ml) 10 mg/ml or 10,000 ppm Selenium RS | Final Volume (ml) |
---|---|---|---|---|
Blank | 50 | 50 | NA | 500.0 |
10 | 50 | 50 | 0.5 | 500.0 |
25 | 50 | 50 | 1.25 | 500.0 |
50 (calibration check standard) | 20 | 20 | 1.0 | 200.0 |
75 | 20 | 20 | 1.5 | 200.0 |
150 | 20 | 20 | 3.0 | 200.0 |
2.7. Instrumental parameters for analysis
ICP-OES conditions
ICP-OES instrument was set up according to parameters listed in table 4 and calibrated with the calibration standards (10, 25, 50, 75, and 150 μg/ml) each time before use. Selenium (Se) concentrations were quantified using external calibration. Initial calibration verification was performed with the calibration check standard (50 μg/ml). Samples were bracketed with the calibration check standard following every sixth sample. The permissible cutoff for quality control (QC) verification was 5.0% of the claimed QC concentration.
Parameters | Settings |
---|---|
Pump Rate | 2.2 ml/min |
Sample Flush Time | 70 sec |
Nebulizer used | Concentric Glass |
Spray Chamber used | Cyclonic Glass |
Center tube | 2.0 mm |
Pump Sample Tube (Tygon) | Color :Orange-White |
Flow Rate of Nebulizer | 0.45 lit./min. |
Flow of Coolant Gas | 12 lit./min. |
Flow of Auxiliary Gas | 0.6 lit./min. |
View | Radial |
RF Power | 1300 watts |
Wavelength | 196.090 nm |
Low Wavelength Integration Time | 15 sec |
High Wavelength Integration Time | 5 sec |
2.8. Validation parameters
2.8.1. System suitability/linearity
Before initiating every validation parameter, all standard aliquots and blank solutions were injected in ICP-OES. The acceptance criterion for relative standard deviation (%RSD) of 5 replicate analyses of the 50 μg/ml selenium calibration checked standard was NMT 2.0%. The calibration check standard was injected regularly to ensure the proper running of ICP-OES. The linearity of the method was proven using the standard aliquot solutions. The correlation coefficient of the 5-point calibration curve was calculated, and the %RSD of the calibration check standard was monitored.
2.8.2. Specificity parameter
Selenium determination was performed on the most sensitive emission line at 196.090 nm. Thus, the specificity was defined as no spectral interferences exhibited at 196. 090 nm, and microwave vessels used for digestion demonstrated no sample carryover. Spectral interferences were evaluated by aspirating blank, calibration standard, and sample solutions into the ICP-OES as well as by collecting full frame profiles of the charge injection device chip in the detector. Method blanks were digested and analyzed to monitor selenium carryover in the microwave vessels.
2.8.3. Method precision
Six samples of Selsun selenium sulfide drug substance were prepared by weighing approximately 0.5 g of drug substance into microwave vessels and digesting according to the procedure. To prepare six samples of Selsun blue dandruff shampoo (SeS2, 1%) with menthol, about 0.5 g of product was weighed into microwave vessels and digested according to the procedure. In addition, six samples of selenium sulfide topical suspension USP (SeS2 2.5%) lotion samples were created through the process of weighing around 0.2 g of product and digesting it according to the same procedure. Selenium concentrations were determined via ICP-OES analysis.
2.8.4. Intermediate precision
Similar to the method precision parameter, six samples of selenium sulfide drug substance, Selsun blue dandruff shampoo (SeS2 1%) with menthol and selenium sulfide topical suspension USP (SeS2 2.5%) lotion samples were prepared and analyzed in ICP-OES on a different day.
2.8.5. Accuracy
Accuracy was performed at three concentrations (each triplicate preparation) ranging from 10 ppm, 50 ppm, and 150 ppm for drug substance and drug product for selenium content. The % recovery in each level was calculated.
2.8.6. Limit of detection and limit of quantitation
The limit of detection (LOD) and limit of quantitation (LOQ) were calculated from linearity solutions. The LOD for selenium was calculated from the linearity curve. Similarly, LOQ was determined.
LOD (µg/ml (ppm)) = 3.3 × Standard deviation/slope
LOQ (µg/ml (ppm)) = 10 × Standard deviation/slope
2.8.7. Robustness
Variations in instrument conditions (Table 5) were performed to evaluate robustness. The highest applied RF power setting for our ICP-OES instrument was 1,350 W. Therefore, the original work plan of 1,500 W was modified to test up to 1,300 W for the robustness study.
Operating Parameters | Settings |
---|---|
Pump Rate (ml./min.) | 1.8, 2.2, 2.6 |
Flow Rate of Nebulizer Gas (lit./min.) | 0.35, 0.45, 0.55 |
Flow Rate of Auxiliary Gas (lit./min.) | 0.5, 0.6, 0.7 |
Applied RF Power (W) | 900, 1100, 1300 |
Integration (sec) | 15, 30, 45 |
3. Results
3.1. Method developmental trials
During solubility studies, the selenium drug substance was found to be insoluble in water. The same result was confirmed in other studies ( 24 ). Further, Selsun blue dandruff shampoo with menthol and topical suspension USP was observed to be water-soluble.
To develop a common sample preparation method for selenium drug substance and drug product, the direct aqueous solution method for sample preparation was disregarded. Thus, the combination of microwave-assisted acid digestion and closed containers was used at higher temperatures without reagent and analyte losses and with a reduction in contamination risks.
In the closed vessel microwave-assisted digestion procedure, different volumes of concentrated nitric acid and hydrochloric acid had no significant impact on the digestion of the sample. Hence, according to Mathew et al. (2017) ( 25 ), molten ammonium hydrogen difluoride was utilized to digest the sample matrix. Ammonium hydrogen difluoride converted metal oxides to fluorides. Upon the addition of concentrated nitric acid, the fluorination reaction continued with the release of hydrofluoric acid (HF).
(NH4F.HF) + HNO3 NH4NO3 + 2 HF
Boric acid was required to complex the remaining hydrofluoric acid.
3.2. System suitability/linearity
The acceptance criterion for the correlation coefficient (r) of the 5-point calibration curve (10-150 μg/ml Se) was NLT 0.999. System suitability was achieved with a correlation coefficient value of 0.999980. (Figure 1 and Tables 6 and 7).
Standard Result (μg/ml) or (ppm) | Standard Result (μg/ml) or (ppm) |
---|---|
0 | 0.0000 |
10 | 9.9336 |
25 | 24.745 |
50 | 50.246 |
75 | 75.473 |
150 | 149.600 |
Correlation coefficient | 0.999980 |
Standard Error | 39.393156315581 |
slope | 101.0963 |
Standard Result (μg/mL) | Standard Result (μg/mL) |
---|---|
50 | 50.78 |
50 | 50.64 |
50 | 50.62 |
50 | 50.50 |
50 | 50.65 |
Average | 50.64 |
Stdev | 0.10 |
%RSD | 0.2 |
3.3 Specificity
The specificity of the method was confirmed by measuring selenium at 196.090 nm, 203.985 nm, and 206.279 nm. Blank, 150 µg/ml calibration standard, 1% SeS2 Selsun blue shampoo, and 2.5% SeS2 Perrigo shampoo samples were measured at the above wavelengths. Hence in specificity solutions, spectral interferences were not observed at 196.090 nm, and selenium was not detected greater than 0.10 µg/ml in method blanks (Figure 2).
3.4 Method precision
The acceptance criterion for %RSD of the 6 samples each for selenium drug substance, Selsun blue dandruff shampoo (SeS2 1%) with menthol, and selenium sulfide topical suspension USP (SeS2 2.5%) was NMT 2.0%. For the drug substance, the %RSD for 6 replicate preparation was 0.2, for shampoo, for 6 replicate preparation, the %RSD was 1.7, and for the topical suspension, for 6 replicate preparation, the %RSD was 1.6 (Table 8).
selenium sulfide – drug substance | selenium content % | selsun blue shampoo (ses2 1%) sample preparation | selenium content % | Perrigo selenium sulfide topical suspension usp (ses2 2.5%) sample preparation | selenium content % |
---|---|---|---|---|---|
1 | 99.5 | 1 | 94.0 | 1 | 100.1 |
2 | 99.2 | 2 | 95.9 | 2 | 101.3 |
3 | 98.9 | 3 | 93.1 | 3 | 100.7 |
4 | 98.9 | 4 | 95.2 | 4 | 102.9 |
5 | 99.1 | 5 | 93.6 | 5 | 99.7 |
6 | 99.3 | 6 | 91.3 | 6 | 103.8 |
Average | 99.2 | Average | 93.8 | Average | 101.4 |
Stdev | 0.23 | Stdev | 1.63 | Stdev | 1.59 |
%RSD | 0.2 | %RSD | 1.7 | %RSD | 1.6 |
3.5 Intermediate Precision
The results of 6 samples prepared and analyzed on different days were compared. The drug substance sample met current monograph specifications with an average selenium content of 99.2% for day 1 and an average of 99.8% for day 2. Both different days met the work plan criteria of NMT 2.0% RSD for 6 samples analyzed. The %RSD of overall precision for 12 samples was 0.6. The shampoo sample met current monograph specifications with an average selenium content of 93.8% for day 1 and an average of 93.6% for day 2. Both different days met the work plan criteria of NMT 2.0% RSD for 6 samples analyzed. The %RSD of overall precision for 12 samples was 1.4. The topical suspension sample met current monograph specifications with an average selenium content of 10.1.4% for day 1 and an average of 100.2% for day 2. Both different days met the work plan criteria of NMT 2.0% RSD for 6 samples analyzed. The %RSD of overall precision for 12 samples was 1.5 (Table 9).
selenium sulfide – drug substance sample preparation | selenium content % | selsun blue shampoo (ses2 1%) sample preparation | selenium content % | perrigo selenium sulfide topical suspension USP (ses2 2.5%) sample preparation | selenium content % | |||
---|---|---|---|---|---|---|---|---|
Day-1 | Day-2 | Day-1 | Day-2 | Day-1 | Day-2 | |||
1 | 99.5 | 98.9 | 1 | 94.0 | 94.1 | 1 | 100.1 | 98.7 |
2 | 99.2 | 98.5 | 2 | 95.9 | 95.5 | 2 | 101.3 | 99.9 |
3 | 98.9 | 100.4 | 3 | 93.1 | 93.3 | 3 | 100.7 | 99.5 |
4 | 98.9 | 100.2 | 4 | 95.2 | 92.8 | 4 | 102.9 | 100.7 |
5 | 99.1 | 99.9 | 5 | 93.6 | 92.8 | 5 | 99.7 | 102.6 |
6 | 99.3 | 100.1 | 6 | 91.3 | 92.9 | 6 | 103.8 | 100.0 |
Average | 99.2 | 99.8 | Average | 93.8 | 93.6 | Average | 101.4 | 100.2 |
Stdev | 0.23 | 0.78 | Stdev | 1.63 | 1.07 | Stdev | 1.59 | 1.30 |
%RSD | 0.2 | 0.8 | %RSD | 1.7 | 1.1 | %RSD | 1.6 | 1.3 |
Overall average | 99.4 | Overall average | 93.7 | Overall average | 100.8 | |||
Overall %RSD | 0.6 | Overall %RSD | 1.4 | Overall %RSD | 1.5 |
3.6 Accuracy
Accuracy is the degree to which observed values and actual values agree. Drug substance recovery rates ranged from 98.5% to 100.7% on average. The average shampoo recoveries fell between 101.6% and 103.0%, which was comparable. For topical suspension, the range of typical recoveries was 99.7% to 101.7% (Table 10).
Sample Name | Sample concentration level | Average Recovery (%) |
---|---|---|
Selenium sulfide | 10 ppm samples | 100.7% |
50 ppm samples | 98.5% | |
150 ppm samples | 99.7% | |
Sample Name | Spiked level | Average Recovery (%) |
Selsun Blue Shampoo (SeS2 1%) | 10 ppm spiked samples | 101.6% |
50 ppm spiked samples | 102.9% | |
150 ppm spiked samples | 103.0% | |
Selenium Sulfide Topical Suspension USP (SeS2 2.5%) | 10 ppm spiked samples | 101.7% |
50 ppm spiked samples | 99.7% | |
150 ppm spiked samples | 101.7% |
3.6 Limit of detection and limit of quantitation
The linearity curve was used to calculate the LOD and LOQ. The LOD values were 1.28 ppm and 3.89 ppm, respectively.
3.7 Robustness
System suitability results of all tested conditions (Figure 3) met acceptance criteria of NMT 2.0% RSD and NLT 0.999 for r.
4. Discussion
Selenium sulfide is commonly used in various dosage forms, such as shampoo and topical suspension, because of its antifungal and antibacterial properties. A common microwave-assisted digestion technique was developed and validated. The precision, specificity, linearity, accuracy, and robustness of the method for estimating selenium in selenium sulfide drug substances and various pharmaceutical dosage forms were demonstrated (Table 11). This newly developed microwave-assisted digestion technique has optimum sensitivity and is highly reproducible and time-saving than the existing methods (Table 12). This method can be applied to numerous matrices for a finished dosage of selenium sulfide formulations.
Parameter | Description | Criteria | Results |
---|---|---|---|
System Suitability Linearity | 5-point calibration curve (10-150 µg/mL selenium). | The correlation coefficient (r) is NLT 0.999. | The correlation coefficient (r) is 0.999980. |
System Suitability Precision | 5 replicate analysis of the 50 µg/mL selenium calibration check standard. | The %RSD is NMT 2.0%. | The %RSD for the 5 continuous readings of Se check standard is 0.2. |
Robustness | Pump rate at 1.8, 2.2, and 2.6 mL/min. | System suitability meets criteria under changes to instrument conditions | All correlation coefficients (r) are greater than 0.999 and all replicate analysis %RSD are NMT 2.0%. |
Nebulizer gas flow at 0.35, 0.45, 0.55 L/min. | |||
Auxiliary gas flow at 0.5, 0.6, 0.7 L/min. | |||
RF power at 900, 1100, 1300 watts. | |||
Integration at 15, 30, 45 sec. | |||
Specificity | Collect full frame profiles of the CID (charge injection device) for blanks, calibration standards and samples. | The emission line at 196.090 nm is free of spectral interferences and microwave vessels used for digestion exhibit no sample carryover. | Spectral interferences are not present at 196.090 nm and Se in methods blanks is not detected greater than 0.10 µg/mL. |
Precision | Six samples of each product under study are prepared and Se concentrations of the digests determined by ICP-OES. | The %RSD of the 6 sample digests is NMT 2.0%. | For the drug substance, the RSD is 0.2%, shampoo, the RSD is 1.7% and for the lotion the RSD is 1.6%. |
Intermediate Precision | For each product under study, results of 6 samples prepared and analyzed by two independent analysts are compared on different days. | Calculated amounts of SeS2 are within current monograph specification of 90.0 – 110.0% of the label claim. The %RSD of the 6 sample digests is NMT 2.0%. %RSD of overall precision on different days for 12 sample digested is NMT 3.0% | The drug substance sample met current monograph specifications with an average selenium content of 99.2% for Day 1 and an average of 99.8% for Day 2. The %RSD of overall precision on different days for 12 sample is 0.6. |
The shampoo sample met current monograph specifications with an average selenium content of 93.8% for Day 1 and an average of 93.6% for Day 2. The %RSD of overall precision on different days for 12 sample is 1.4. | |||
The topical suspension sample met current monograph specifications with an average selenium content of 10.1.4% for Day 1 and an average of 100.2% for Day 2. The %RSD of overall precision on different days for 12 sample is 1.5 | |||
Accuracy | Nine accuracy samples (triplicate preparation at sample concentration of 10 ppm, 50 ppm and 150ppm for drug substance. | Recovery of the spikes is between 95.0 – 105.0%. | The average spike recovery of drug substance was 100.7%, 98.5% and 99.7% at three different levels. |
Nine spiked Samples each for shampoo and lotion were prepared by spiking selenium standard at 10 ppm, 50 ppm and 150ppm levels | The average spike recovery of shampoo was 101.6%, 102.9% and 103.0% at three different levels. | ||
The average spike recovery of Topical suspension was 101.7%, 99.7% and 101.7% at three different levels. | |||
Limit of detection (LOD) & limit of quantitation (LOQ) | The linearity curve was used to calculate the limits of detection (LOD) and quantitation (LOQ). | Report the results | Limit of detection was calculated form the linearity curve. The limit of detection was 1.28 ppm and 3.89 ppm respectively. |
Instrumental Technique | Method Details | Merits | Demerits | References |
---|---|---|---|---|
voltammetric analysis | Complexing Selenium with complexing agents (like selenocystine and dimethyldiselenide or (iron (III)-loaded) followed by extraction with agents like with dichloromethane or 0.5 ml HCl 37% by heating or | Cost effective | Low sensitivity Tedious multistep process and highly toxic. | 4 , 5 |
No method is validated | ||||
spectrophotometry | Chemical modification using agents like 4-aminoantipyrine (4-amino-1,2-dihydro-1,5-dimethyl-2-phenyl-3H-pyrazole-3-one; 4-AAP) by selenium in presence of acidic medium and the coupling with N-(naphthalen-1-yl)ethane-1,2-diamine dihydrochloride (NEDA) to give a violet color derivative or reacting selenium with 2,3-diaminonaphthalene was reinvestigated with bromide ion as a catalyst or | Optimum recovery | Low sensitivity due to cooling step. Few methods are only validated | 6 , 7 |
chromatography (HPLC) | pre-column derivatization with 2,3-diaminonaphthalene or sodium salt of n-octanesulfonic acid as ion-pairing modifier in column | Good recovery | Complex multistep procedure. High skill required | 8 , 9 |
atomic absorption spectroscopy (AAS) | MWD Digestion followed by Hydride generation | Optimum recovery and sensitivity | Time consuming and toxic sample preparation technique | 10 - 14 |
atomic fluorescence spectroscopy | 2,3-diaminonaphthalene (DAN) complexing or Se-2,3-diaminonaphthalene complex formation | Optimum sensitivity | Time consuming process. Low reproducibility | 15 - 17 |
ICP-OES and ICP-MS | Instrumental Neutron Activation Analysis (INAA) with on-line digestion of the fraction, reduction and hydride formation or complexation with surfactant of p-octyl polyethyleneglycolphenyl ether (Triton X-100) | High sensitivity. High precision | High skill required. Costly | 18 - 20 |
Inhouse newly developed method | Direct Microwave digestion followed by ICP-OES | Simple sample preparation method, less toxic, low skill required and highly reproducible | Optimum sensitivity (less sensitive compared to ICP-MS and GFAAS and more sensitive than spectrometric, Volta metric and chromatograph and AAS technique) | Not Applicable |
Acknowledgment
The authors are thankful to the management of GITAM (Deemed to be University), Visakhapatnam, Andhra Pradesh, India.
Authors' Contribution
Conceptualization, methodology and original draft preparation: A.K.P. Supervision, reviewing and editing: R.S.
Ethics
We hereby declare all ethical standards have been respected in preparation of the submitted article.
Conflict of Interest
The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
References
- Alexander J. Handbook on the Toxicology of Metals (Fifth Edition). Editors: Gunnar F, Nordberg, Max Costa. Academic Press. 2022; 729-771.
- International Council for Harmonization of Technical Requirements for Pharmaceuticals for Human Use. ICH harmonized guideline: guideline for elemental impurities – Q3D (R1), European Medicines Agency, Amsterdam. 2019.
- Soruraddin MH, Heydari R, Puladvand M, Zahedi MM. A new spectrophotometric method for determination of selenium in cosmetic and pharmaceutical preparations after preconcentration with cloud point extraction. Int J Anal Chem. 2011;729651.
- Ramadan AA, Mandil H, Ozoun A. Determination of Se (IV) in Pharmaceuticals by Using Different Operating Modes of Pulse Anodic Stripping Voltammetric Analysis with Methylene Blue-Nafion Modified Gold Electrode. Asian J. Chem. 2013; 25(6)
- Stoica AI, Babaua GR, Iorgulescu EE, Marinescu D, Baiulescu GE. Differential pulse cathodic stripping voltammetric determination of selenium in pharmaceutical products. J Pharm Biomed Anal. 2002; 30(4):1425-9.
- Kumar KS, Suvardhan K, Kang SH. Facile and sensitive determination of selenium (IV) in pharmaceutical formulations by flow injection spectrophotometry. J Pharm Sci. 2008; 97(5):1927-33.
- Sanz J, Gallarta F, Galbán J, Castillo JR. Determination of selenium by hydride generation ultraviolet-visible molecular absorption spectrometry with diode-array detection. Analyst. 1988; 113(9):1387-91.
- Ma Y, Xiang F, Jin W, Liao N, Yu L. Selenium accumulation in mycelia of Flammulina velutipes during fermentation determined by RP-HPLC. Zeitschrift für Naturforschung C. 2009; 64(5-6):382-6.
- Kulikov AU. Determination of selenium (IV) in pharmaceuticals and premixes by micellar liquid chromatography. J. Pharm. Biomed. Anal. 2007; 43(4):1283-9.
- Kazi TG, Kolachi NF, Afridi HI, Brahman KD, Shah F. Determination of total selenium in pharmaceutical and herbal supplements by hydride generation and graphite furnace atomic absorption spectrometry. J. AOAC Int. 2014; 97(6):1696-700.
- Gamiz-Gracia L, De Castro ML. Determination of selenium in nutritional supplements and shampoos by flow injection-hydride generation-atomic fluorescence spectrometry. Talanta. 1999; 50(4):875-80.
- Lavi N, Mantel M, Alfassi ZB. Determination of selenium in biological materials by neutron activation analysis. Analyst. 1988; 113(12):1855-9.
- Chan CC, Sadana RS. Determination of arsenic and selenium in environmental samples by flow-injection hydride generation atomic absorption spectrometry. Anal. Chim. Acta. 1992; 270(1):231-8.
- Bujdoš M, Kubová J, Streško V. Problems of selenium fractionation in soils rich in organic matter. Anal. Chim. Acta. 2000; 408(1-2):103-9.
- García JB, Krachler M, Chen B, Shotyk W. Improved determination of selenium in plant and peat samples using hydride generation-atomic fluorescence spectrometry (HG-AFS). Anal. Chim. Acta. 2005; 534(2):255-61.
- Smrkolj P, Stibilj V. Determination of selenium in vegetables by hydride generation atomic fluorescence spectrometry. Anal. Chim. Acta. 2004; 512(1):11-7.
- Sabé R, Rubio R, Garcı́a-Beltrán L. Selenium determination in urine with atomic fluorescence detection. Anal. Chim. Acta. 2001; 436(2):215-21.
- De Brätter VN, Recknagel S, Gawlik D. Speciation of Se, Fe and Zn in human milk whey: the use of instrumental neutron activation analysis (INAA) to corroborate element profiles measured with inductively coupled plasma atomic emission spectrometry (ICP-AES). Fresenius J. Anal. Chem. 1995; 353:137-42.
- Wen S, Zhu X, Wei Y, Wu S. Cloud point extraction-inductively coupled plasma mass spectrometry for separation/analysis of aqueous-exchangeable and unaqueous-exchangeable selenium in tea samples. Food Anal. Methods. 2013; 6:506-11.
- Bakirdere S, Volkan M, Ataman OY. Speciation of selenium in supplements by high performance liquid chromatography-inductively coupled plasma-mass spectrometry. Anal. Lett. 2015; 48(9):1511-23.
- Mazej D, Falnoga I, Veber M, Stibilj V. Determination of selenium species in plant leaves by HPLC–UV–HG-AFS. Talanta. 2006; 68(3):558-68.
- USP Selenium Sulfide. In: USP–NF. Rockville, MD: USP. 2022. DOI
- USP Selenium Sulfide. Topical Suspension In: USP–NF. Rockville, MD: USP. 2022. DOI
- National Center for Biotechnology Information. PubChem Annotation Record for SELENIUM, ELEMENTAL, Source: Hazardous Substances Data Bank (HSDB). 2023.
- O'Hara MJ, Kellogg CM, Parker CM, Morrison SS, Corbey JF, Grate JW. Decomposition of diverse solid inorganic matrices with molten ammonium bifluoride salt for constituent elemental analysis. Chem Geol. 2017; 466:341-51.